Flow Computer:                                                                                                            home
Model FC200


DESCRIPTION


Flow Systems introduces the lightweight and mobile Flow Computer model FC200.  The FC200 offers a greater economy than its FC500 counterpart, and is designed for applications requiring fewer features and a lower price.  The FC200 interfaces with single, manually interchangeable, Sonic Nozzles and Sub-Sonic Venturis.  Multiple levels of accuracy are available with highly repeatable results. 



                                      Flow Computer - FC200-VEN
                                       Base Unit Dimensions: 
                                     2.5 in. H x 8 in. W x 12 in. D

FEATURES

»  MS Windows™-based Setup Software and Interface allow Easy Configuration and Monitoring Capabilities.

»
  USB Communication.

»  PC Connection Required. (No Stand-Alone Operation.)

»  Self-Contained Package including Instrumentation.

»  Calibration Software to Maintain Internal Instruments.

»  Test Set-up Screen.

»  Flowmeter Configuration.

»  Optional Internet Support and Diagnostics.

»
  Data Logging Capabilities via USB Interface.

»  Best Economy.

 


SOLUTION

This unique solution (hardware and software) combines Instrumentation, Data Acquisition and Control, Computation, Monitoring and communication with a local, customer-supplied personal computer or laptop.   Accurate Measurement via User-Friendly Controller and High System Availability are key features of this test instrument.

The data acquisition, control subsystems and all instrumentation are standard, readily available Commercial Off-The-Shelf (COTS) items. This ensures that servicing or replacement of any item can be achieved in a timely manner, providing high-availability for the test system.

The FC200 works with single Sonic Nozzles and Sub-Sonic Venturis. For those users requiring increased accuracy or advanced functionality, we suggest the FC500 model flow computer.  




  Flow Computer - FC200-CFV

 

Flow Computer - FC200-CFV with interchangeable nozzles and meter runs

 


 

 


Configuration / Monitoring Software:

The FC200 software was developed using National Instruments™ LabVIEW™ and uses a USB connection to communicate with the FC200. This package gives the user the following functionality:

 


Test Set-up Screen:

Configures the FC200 for meter type, fluid, and units.

Real Gas Properties for most Gasses.



Flowmeter Configuration:



Calculations for Sonic Nozzles are in accordance with ASME MFC-7M / ISO 9300. 
Computations for Sub-Sonic Venturis meet the requirements of ASME MFC-3M / ISO 5167.

Iterative Solution for Discharge Coefficient (Cd) based on Actual Calibration Data or Theoretical Models.

 

 

Data Monitoring and Logging:

Data may be logged automatically at a user-defined interval.

Flow Totalizing / Batching: Allows the User to Start, Stop, Clear, and Log Batch Data.


 

Calibration Assistant for Maintenance of Internal Instrumentation:


Optional Internet Support and Diagnostics:


Allows FLOW SYSTEMS to observe and / or control a system remotely for training, optimization, or troubleshooting purposes. 
This service requires a temporary, high-speed data line.

 

SPECIFICATIONS:

 

Model Code Meter Type Supported
FC200-CFV Sonic Nozzle
FC200-VEN Venturi

 

Options
   -01        Humidity Sensor (Atmospheric Air)
   -02        Dew Point Sensor (Compressed Air)

 

Accuracy

The FC200 uses a 16-bit data acquisition system to measure all analog inputs. The following tables state percent of reading and mass flow accuracy for both available meter types and corresponding instrumentation configurations.

 

Sonic Nozzles

Configuration Mass Flow Accuracy Comments
Level 1 +/-6.45% of Reading Assumes: diameter 0.031" uncalibrated nozzle, 6:1 pressure range (+/-0.1% full scale) & 40° to 100°F gas temperature.
Level 1 +/-0.88% of Reading Assumes: diameter 0.50" uncalibrated nozzle, 6:1 pressure range (+/-0.1% full scale) & 40° to 100°F gas temperature.
Level 2 +/-0.56% of Reading Assumes: +/-0.50% CEESI calibration, 6:1 pressure range (+/-0.04% full scale) & 40° to 100°F gas temperature.
Level 3 +/-0.29% of Reading Assumes: +/-0.25% CEESI calibration, 6:1 pressure range (+/-0.02% full scale) & 40° to 100°F gas temperature.

An optional humidity sensor is available for moist air applications.

 

 

 

Sub-Sonic Venturis

Configuration Mass Flow Accuracy

Comments

Level 1 +/-1.41% of Reading Assumes: Uncalibrated Venturi, 6:1 pressure range (+/-0.1% full scale), 10:1 Dp range (+/-0.3% of reading) & 40° to 100°F gas temperature.
Level 2 +/-0.61% of Reading Assumes: +/-0.5% CEESI calibration, 6:1 pressure range (+/-0.04% full scale), 10:1 Dp range (+/-0.3% of reading) & 40° to 100°F gas temperature.
Level 3 +/-0.43% of Reading Assumes: +/-0.25% CEESI calibration, 6:1 pressure range (+/-0.02% full scale), 10:1 Dp range (+/-0.3% of reading) & 40° to 100°F gas temperature.

An optional humidity sensor is available for moist air applications.

 

 

Connections - Bench Top or Rack Mount:



Power Requirements: 100-240 VAC /  50 - 60Hz / 1 Phase  / 1 AMP

Communication:

USB

 

home