Test Set-up Screen:
Configures the
FC175 for meter type, fluid, and units.
Real
Gas Properties for most Gasses.
Flowmeter Configuration:
Calculations for Sonic Nozzles are in accordance with ASME
MFC-7M / ISO 9300.
Computations for Sub-Sonic Venturis meet the requirements of ASME MFC-3M / ISO 5167.
Iterative
Solution for Discharge Coefficient (Cd)
based on Actual Calibration Data or
Theoretical Models.
Data Monitoring and Logging:

Data
may be logged automatically at a
user-defined interval.
Flow Totalizing /
Batching: Allows the User to Start, Stop, Clear, and Log Batch
Data.
Calibration Assistant for Maintenance of Internal Instrumentation:

Optional
Internet Support and Diagnostics:
Allows FLOW SYSTEMS to observe and / or control a system remotely for training, optimization, or troubleshooting purposes.
This service requires a temporary, high-speed data line.
SPECIFICATIONS:
| Model
Code |
Meter
Type Supported |
| FC175-CFV |
Sonic
Nozzle |
| FC175-VEN |
Venturi |
The
above come with a Calibrated Thermister
Probe for Temperature Measurement.
Optional
Absolute Pressure Transmitters for Sonic
Nozzle Application:
| Option
Code |
Description |
| -01 |
Level
1 with P2 Gauge |
| -01A |
Level
1 with P2 Transmitter |
| -02 |
Level
2 with P2 Transmitter |
| -03 |
Level
3 with P2 Transmitter |
| -04 |
Level
4 with P2 Transmitter |
| -05 |
Level
5 with P2 Transmitter |
Optional
Absolute Pressure and Differential Pressure
Transmitters
for Sub Sonic Venturi Application:
| Option
Code |
Description |
| -05 |
Level
1 Transmitters (2) |
| -06 |
Level
2 Transmitters (2) |
Humidity:
| Option
Code |
Description |
| -07 |
Humidity
Sensor (Atmospheric Air) |
| -08 |
Dew
Point Sensor (Compressed Air) |
Connections
- Bench Top
| Power
Requirements: |
Wall
Power Supply Included |
| Communication: |
USB |
Accuracy
The FC175 uses a 15-bit data acquisition system to measure all analog inputs. The following tables
state percent of reading and mass flow accuracy for both available meter types and corresponding instrumentation configurations.
Sonic Nozzles
| Options |
Configuration |
Mass Flow Accuracy |
Comments |
-01,
-01A
|
Level 1 |
+/0.95% of Reading |
Assumes:
+/-0.50%
CEESI
calibration, 4:1 pressure range
(+/-0.2% full scale) and 40° to 100°F gas temperature.
(+/-0.2%). |
|
-02
|
Level 2
|
+/-0.65% of Reading |
Assumes:
+/-0.50% CEESI calibration, 4:1 pressure range
(+/-0.1% full scale) and 40° to 100°F gas temperature.
(+/-0.2%). |
|
-03
|
Level 3
|
+/-0.54% of Reading |
Assumes:
+/-0.50% CEESI calibration, 4:1 pressure range
(+/-0.05% full scale) and 40° to 100°F gas temperature.
(+/-0.2%). |
|
-04
|
Level 4
|
+/-0.27% of Reading |
Assumes: +/-0.25% CEESI calibration,
4:1 pressure range (+/-0.02% full scale)
and 40° to 100°F gas temperature.
(+/-0.2%). |
|
|
An optional
humidity sensor is available for moist air applications.
|
Sub-Sonic Venturis
| Options |
Configuration |
Mass Flow
Accuracy |
Comments
|
|
-05
|
Level 1 |
+/-0.61% of Reading |
Assumes: Uncalibrated
Venturi, 4:1 pressure range
(+/-0.05% full scale), 10:1 Dp range
(+/-0.02% of full scale) and 40° to 100°F gas temperature.
(+/-0.2%). |
|
-06
|
Level 2 |
+/-0.43% of Reading |
Assumes: +/-0.5%
CEESI calibration,
4:1 pressure range (+/-0.02% full scale), 10:1 Dp range
(+/-0.01% of full scale) and 40° to 100°F gas temperature.
(+/-0.2%). |
|
|
An optional